Accuracy
• See the Specifications section for operating conditions.
• Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount)
at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 75% relative humidity.
• Accuracy is for the CMA-60 only. See the instruction manual supplied with your multimeter
for its accuracy.
Accuracy Table
R
n a
e g
0 1
A 0
0 6
A 0
* Accuracy is specified for DC and 45 to 400 Hz AC.
Specifications
Measurement Range: 0A to 600A
Voltage Output:
100 Amp Range: 10 mV output per 1 amp input
600 Amp Range: 1 mV output per 1 amp input
Output Impedance: 100 kΩ
Temperature Coefficient: 0.2 x (specified accuracy) per °C below 18 °C or above 28 °C
Operating Conditions: 0 °C to 50 °C (32 °F to 122 °F), 0% to 75% relative humidity
Elevation: 2000 m (6500') maximum
Indoor use only
Storage Conditions: –20 °C to 60 °C (–4 °F to 140 °F), 0% to 80% relative humidity.
Remove battery.
Jaw Opening: 35 mm (1.38")
Battery: 9-Volt battery (NEDA 1604, JIS 006P or IEC 6LF22)
Pollution Degree: 2
Overvoltage Category: Category II, 600 Volts
8
I
p n
t u
A
0
o t
0 1
A 0
±
1
0 0
o t
0 4
A 0
±
4
0 0
o t
0 6
A 0
±
c c
r u
c a
* y
2 (
%
+
2
) A
2 (
%
+
2
) A
3 (
%
+
2
) A