Sterilizer Type
Configuration
Gravity
Double Wrapped N/A
Displacement
Pre-vacuum
Double Wrapped 4 pulses
Examine all instruments prior to use. If particles are visible, repeat the cleaning and sterilization process.
Preconditioning
Minimum
Temperature
132-137 °C
(270-279 °F)
132-137 °C
(270-279 °F)
3
Exposure Time
Minimum Dry Time
15-18 minutes
40 minutes
3-18 minutes
30 minutes