Technical Description Concerning Electromagnetic Emission
Table 1: Declaration - electromagnetic emissions
Guidance and manufacturer's declaration – electromagnetic immunity
The model TL30 is intended for use in the electromagnetic environment specified below.
The customer or the user should assure that It is used in such an environment.
Immunity test
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
lEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV line to line
lEC 61000-4-5
±2 kV line to earth
Voltage dips,
<5 % U
short interrup-
for 0.5 cycle
tions and vol-
40 % U
tage variations
for 5 cycles
on power supp-
70% U
ly input lines
for 25 cycles
IEC 61000-4-11
<5% U
for 5 sec
Power frequen-
3 A/m
cy (50/60 Hz)
magnetic field
lEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601
test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±2 kV line to earth
(>95% dip in U
)
<5 % U
T
T
for 0.5 cycle
(60% dip in U
)
40 % U
T
T
for 5 cycles
(30% dip in U
)
70% U
T
T
for 25 cycles
(>95 % dip in U
)
<5% U
T
T
for 5 sec
Not applicable
Compliance
level
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
(>95% dip in U
)
Mains power quality should be that
T
T
of a typical commercial or hospital
(60% dip in U
)
environment. If the user of the
T
T
model TL30 require continued ope-
(30% dip in U
)
ration during power mains interrup-
T
T
tions, it is recommended that the
(>95 % dip in U
)
model 168 DAYLUX Therapy Light
T
T
be powered from an uninterruptible
power supply or a battery.
Not applicable
64
Electromagnetic environment –
guidance