Microplate Shaker
1.4. Installation
1.
Important: choose the space on your bench where the Ohaus High Speed Microplate Shaker will stay.
Because of the amount of energy dissipated, the unit is held onto your lab bench with 12 rubber feet
(suction cups). The unit is VERY difficult to remove from work surface and should be done with caution.
Use paper under the feet to keep them from attaching to the bench while determining the final location.
When final location is determined, remove the paper.
2.
Be sure the power switch is in the off position.
3.
The Ohaus High Speed Microplate Shaker comes with a connector plug that is inserted into the IEC
connector on the back of the unit first, and then it can be plugged into a properly grounded outlet. The
120v unit plugs into a 120 volt, 50/60 Hz source. The 230v unit plugs into a 230 volt, 50/60 Hz source.
4.
Your Ohaus High Speed Microplate Shaker is ready for use.
1.5. Overview
Model:
Controls:
Overall dimensions (L x W x H):
Tray dimensions (L x W ):
Power:
Power consumption:
Fuses:
Speed range:
Timer:
Orbit:
Net Weight:
Ship weight:
Capacity:
Communication:
1.888.610.7664
SHHSMPDG
See section 1.6
15.5 x 12 x 12.8" (39.4 x 30.5 x 32.5cm)
12.25 x 11" (31.1 x 27.9cm)
120V, 50/60Hz / 230V, 50/60Hz
120W
5mm x 20mm, 5 amp quick acting (2 required)
600 to 2500rpm
1 second to 160 hrs
.014" (3.6mm)
48.1 lb / 21.8 kg
50lbs / 22.7kg
7 lbs / 3.2 kg or 48 micropl;ates in stacks up to 5" / 12.7
cm high
RS-232 Interface
www.calcert.com
15
10
20
5
25
0
30
EN-3
sales@calcert.com