Télécharger Imprimer la page

Thermo Fisher Scientific Isotemp Serie Installation Et Fonctionnement page 26

Réfrigérateurs de chromatographie et à usage général

Publicité

Les langues disponibles

Les langues disponibles

Maintenance
Cleaning the Cabinet
Interior
WARNING: Disconnect equipment from main
power before attempting any maintenance to
equipment or its controls unless stated
otherwise.
To clean the cabinet interior, remove the shelves, use a
solution of water and a mild detergent for cleaning. Rinse the
interior storage components and wipe them dry with a soft
cloth.
Cleaning the Unit
(Chromatography
Refrigerators)
Use a solution of water and a mild detergent for cleaning.
Lightly spray the interior storage components and wipe them
dry with a soft cloth or spray the cloth first and then wipe
interior surfaces.
Do not spray or pour directly into holes and gaps in the
chromatography safety circuit assembly for cleaning. Use a
wet wipe to clean around sensor box and outlet.
If the sensor led turns red after cleaning, manually reset the
safety circuit assembly using the manual reset switch on the
right side of header panel.
Cleaning the Condenser
CAUTION: Condensers should be cleaned at
least every six months; more often if the
laboratory area is dusty. In heavy traffic areas,
condensers load with dirt more quickly. Failure to
keep the condenser clean can result in
equipment warm-up or erratic temperatures.
CAUTION: Never clean around the condensers
with your fingers. Some surfaces are sharp.
The condenser is located at the bottom in front of the unit.
General Purpose and Chromatography Refrigerators
To clean the condenser, complete the following steps:
1. Disconnect the power.
2. Vacuum the condenser and clean up any loose dust.
3. Reconnect power.
Automatic Defrost
The defrosting process on all the models is primarily
accomplished by air circulated during off-cycle. This heat free
process ensures that the temperature is not affected by the
defrost cycle. Defrost cycle works on adaptive defrost
method which is both temperature and time controlled.
When the evaporator sensor reaches below 0°C, the defrost
cycle will be initiated. Initially the defrost cycle continues for a
minimum duration of 5 minutes and after that the defrost
cycle will be terminated once the evaporator sensor
temperature reaches 4°C. In case the evaporator sensor
doesn't reach the temperature of 4°C within 30 minutes from
the start of defrost cycle, the defrost cycle will be terminated.
Gasket Maintenance
Periodically check the gaskets around the door for punctures
or tears. Leaks are indicated by condensation or frost which
form at the point of gasket failure. Make sure that the cabinet
is level.
Keep the door gaskets clean and frost free by wiping gently
with a soft cloth.
To check the door seal, complete the following steps:
1. Open the door
2. Insert a strip of paper (a couple of inches wide) between
the door gasket and the cabinet flange and close the
door.
3. Slowly pull the paper strip from the outside. You should
feel some resistance.
4. Repeat this test at 4-inch intervals around the door. If the
door does no seal properly, replace the gasket.
Maintenance |
21

Publicité

loading