edge of the unit to support the pallet. Once in position,
gently roll the unit off of the pallet.
With Forklift
Required Tools for Single Door Unit
(12, 25 & 30 cu.ft. Models): Phillips head screwdriver, 7/16",
5/16"-18" wrenches, wrench supplied and a minimum of
2 people.
Required Tools for Double and Triple Door Unit
(45, 49 & 72 cu.ft. Models): Phillips head screwdriver, 7/16",
5/16"-18" wrenches and a minimum of 2 people.
All the units are shipped with factory installed feet with rolling
casters supplied inside of the unit. The following instructions
demonstrate how to remove the cabinet from its shipping
base (pallet) and install the casters without use of a forklift.
1. Remove the screws, bolts and pallet by following the
instructions given in section Without Forklift section
from step 1 to step 3.
Note: Always use caution when lifting unit with forklift to
ensure not to damage or drop unit.
2. Carefully position forks to lift unit from rear. Ensure unit is
fully positioned on fork truck prior to raising the forks.
3. Raise the unit approximately six inches to allow access to
the feet and casters and pull out the pallet.
Person 1: Stand at the front of the unit and provide
additional support to the unit.
Person 2: If feet are desired, lower all 4 factory installed
10 |
Installation
feet (using the 3/8" wrench to loosen if necessary). If
rolling casters are desired, remove the feet and install the
4 casters and tighten using wrench. Refer to Without
Forklift section (step 5, 6 and 7) for tightening the
casters.
Chromatography
Refrigerators
The chromatography refrigerator is equipped with a safety
circuit. When the safety circuit is tripped, the safety sensor
senses a flammable gas and the power to the internal outlet
will be discontinued. The user will need to manually reset the
safety circuit assembly using the reset switch at the right side
of header panel before the power is restored. To do this,
toggle the switch located on the top right side of the header
panel beside the remote alarm contacts. There will be a
power delay to the outlet for approximately 4 minutes where
the red LED will be illuminated.
When there is power at the outlet, the red LED will turn off and
the green LED illuminates.
If the circuit trips from RED to GREEN repeatedly or never
turns GREEN, contact Customer Service.
Flammable materials should not be stored in the refrigerator.
Lights
The GFCI outlet assembly is located inside the unit on the rear
wall at the center.
The outlet assembly has two LEDs:
•
RED (on the top) - When the red LED is ON, it indicates
there is no power to the AC outlet inside the unit.
•
GREEN (on the bottom) - When the green LED is ON, it
indicates there is power to the AC outlet inside the unit.
General Purpose and Chromatography Refrigerators