8.1.
Messages and fault codes
0
Analysis cell connection fault
Calibration required
1
2
Replace the filter
3
Replace sensor
4
Pre-heating fault
5
Air sensor fault
Calibration fault
6
Calibration fault
7
8
Analysis fault
9
USB key missing
10
Update required
11
Update script cannot be read
12
USB key missing
9. History
On the welcome page, the "History" button allows you to see the previous tests carried out using the
GASFIND analyser.
Press "Previous" and "Next" to see the different sets of test results.
Press "Print" to print the corresponding test results.
Press "Return" to return to the welcome page.
810G71 - Version V2 – Janvier 2022
Contact SNDC
Press "Calibration"
Change the filter on the front and press "Filter replaced" in the
"Settings" menu
Contact SNDC
Contact SNDC
Contact SNDC
Place the equipment in the open air and check that the test hose
has been correctly disconnected from the ECS circuit. Press
"Calibration".
If the problem persists, contact SNDC.
Place the equipment in the open air and check that the test hose
has been correctly disconnected from the ECS circuit. Press
"Calibration".
If the problem persists, contact SNDC.
Check that the test hose has been correctly connected to the
ECS circuit and that the coupler valve is open.
Insert a USB key
Copy the update file onto the USB key
The update file is corrupt. Contact SNDC
Insert a USB key