Télécharger Imprimer la page

SNDC ECOCLIM GASFIND R134A Notice D'utilisation page 28

Analyseur de réfrigérant

Publicité

Les langues disponibles

Les langues disponibles

8.3.
Results
Each result will be assigned a test number.
The results window will display the composition of the sample analysed:
R134a
-
R12/R1234yf
-
R22
-
Hydrocarbons
-
Unknown refrigerant
-
Air
-
The percentages of unknown refrigerant and air are recorded separately. Please note that the sum of the
percentages may be greater than 100%.
Press the "Print" button to print out the test results.
Press the "Return" button to return to the welcome page.
If the percentage of the refrigerant being tested (R134a or R1234yf) is greater than or equal to 98%, the
results window will be green. If the percentage of the refrigerant being tested is between 95% and 98%,
the window will be orange. If the percentage of the refrigerant being tested is less than 95%, the window
will be red.
810G71 - Version V2 – Janvier 2022
P. 28
Traduction de la notice originale

Publicité

loading

Ce manuel est également adapté pour:

Gasfind 1234yf