All manuals and user guides at all-guides.com
EN-34
5.2 Replacing Power Fuse
The use of a fuse of a different type or with a different value, or
bridging or shunting the fuse is not allowed and can possibly
cause a hazard to your safety and lead to instrument damage!
To replace the fuse:
1.
Put the tip of a screwdriver or a test pen in the circled position below, and then
pull the fuse holder out.
2.
Take out the old fuse and then push the new fuse into the slot. Choose the
correct fuse to replace according to the fuse specification listed in the
Dimensions Section.
Note: the unit comes with a spare fuse in the fuse holder (in the circled
position above).
Attention: If the fuse is good and power is available at the outlet, the cord or
unit may be defective. Try a new cord. If this does not work, the unit should be
sent back for servicing.
3.
Push the fuse holder back in to finish the replacement.
WARNING: Electric Shock Hazard. Disconnect
the equipment from the power supply before
replacing the fuse.
TM
Guardian
Hotplate-Stirrers