ProMed 620 Instructions D'utilisation page 12

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 18
I
nFormatIon on the unIt
For disinfection of surfaces and devices:
Promed Pure FD (Product No. 330815)
For disinfection and cleaning of instruments and grinding heads:
Promed UC-50 (Product No. 330210)
• If needed, the power adapter (Product No. 290426) and the handpiece (Product No. 200601)
can be purchased from Promed separately.
o
PeratInG the unIt
Before you switch the unit on, you must set the rotating speed (rotary knob (G))
to zero.
A. Insert the power adapter (A) into the power socket and the power adapter plug (B), on the
back of the device, into the provided power adapter socket (C).
B. Insert the plug of the handpiece into the socket (E) on the right side of the device.
C. Select a suitable grinding head. Insert the grinding head into the grinding head holder (D) at the
DE
front of the handpiece. The mechanism locks automatically as soon as the grinding head starts to
rotate. Never use force to remove or insert a grinding head! Grinding heads with a shaft diameter of
2.332 mm to 2.350 mm (0.091 to 0.093 inches) fit exactly into the grinding head holder without
EN
you having to use excessive force. To remove a grinding head, switch off the device with the
rotation switch (G) (middle position), making sure that the rotation switch (G) clearly locks into
place. Now you can remove the grinding head by simply pulling it out.
FR
D. Control of the rotation direction and speed:
1. Selection of the rotation direction:
You make this selection by turning the rotation switch (G) to the left or right.
IT
Counterclockwise: the device rotates backwards.
Clockwise: the device rotates forward.
2. Selection of speed:
ES
The speed is regulated by turning the rotation switch (G) to the left or to the right. The more you
turn, the higher the rotationspeed. The selected speed is displayed by the speed indicator (H).
3. Indication of speed:
The speed indicator (H) is to the right of the rotation switch (G). It displays the selected speed
NL
using 5 LED lights. If all lights are flashing, the device runs at the highest speed.
E. Before the first use, make sure that you let the handpiece run for about 30 minutes at approx.
8 000 rpm (2 LED lights of the speed indicator (H) should flash). This is necessary to bed in the
RU
carbon brushes of the motor.
Note:
PL
A different running noise between clockwise and counterclockwise rotation is normal and
caused by bedding in of the carbon brushes or by running in of bearings.
SE
FI
/