IDENTIFICATION OF THE MEDICAL EQUIPMENT
Category:
Brand/model/type
Serial number:
Date of manufacture:
Availability and condition of the instruction manual, loading plate, identification plate/labelling/printing
Good general condition (structure, base, articulations, welds, attachments, etc...). Absence of rusting
Good general cleanliness
Good stability of the hoist and good operation of the base opening/closing
Checking of the tightness of screws (suspension pin on the lifting arm, weighing module attachments,
etc.), absence of play.
Good condition of slings and cradle or hammock, including the attachment system
Good rotation of the gripping bar and good condition of the hooks
Good operation of the wheels and brakes (swivelling, running, locking)
Good lubrication of articulations (absence of noise nuisance, squeaking and hard points, etc.)
Good condition of electric cables, control boxes, connectors and actuators (jacks, motors, etc.)
Good condition of controls and indicator lights
Locking out of operational functions and correct operation of visual and audible alarms
Emergency stop operates correctly
Battery and charger operate correctly
Obstacle stop operates correctly
Correct operation of the emergency descent
Test at maximum load (1 complete cycle )
INSPECTION OF THE AMPLITUDE OF MOVEMENTS
Satisfactory operation of the jack (elevation/descent)
Maximum and minimum height (according to the manufacturers specifications)
Tester model:
Continuity :
Breakdown:
COMMENTS:
Dated:
Name:
NA: Not applicable-C: Conform-NC: Not Conform
Febrero 2022
INSPECTION ANNUAL OF PATIENT HOISTS
Inventory number:
VISUAL INSPECTION (according to EN ISO 10535)
SAFETY INSPECTION (according to EN ISO 10535)
ELECTRICAL SAFETY (according to EN 60601-1)
Inspector's signature and stamp
L0900/ UDI : 8436571820464 WINNCARE
Department/Location:
Serial number:
580154
ESTABLISHMENT
NA
NA
NA
NA
Date of the next inspection