Technical specifications
Measuring instructions
|
For optimal measuring precision: position the cable precisely at the
point inside the clamping jaws where the marker lines intersect.
|
If the cable is positioned improperly, the measurement error will be
no greater than 1.5%.
Accuracy: (% of measured reading + number of digits) at 18 °C to
|
28 °C (64 °F to 74°F) and relative humidity < 75%.
AC current measurements (True RMS)
Range
40 A
200 A
DC current measurements
Range
40 A
200 A
(Max: 300 ADC)
Resolution
0.01 A
0.1 A
Resolution
0.01 A
0.1 A
34
Tolerance
± 3.0 % +5 digits
Tolerance
± 2.5 % +8 digits