3.2.2.22. Mode.Parameter.StopCond.VStop.Type
Mode.Parameter.StopCond.VStop.V
Mode.Parameter.StopCond.VStop.Factor
Stop volume.
If an absolute stop volume (abs.) has been selected, the volume in mL is
valid.
A relative stop volume (rel.) is dispensed as a function of the sample size:
Stop volume in mL = smpl size ∗ factor
The factor is valid.
OFF means that the criterion is not monitored.
3.2.2.23. Mode.Parameter.StopCond.FillRate
Filling rate in the titration in mL/min. Max. means maximum possible filling
rate with the Exchange Unit in current use.
3.2.2.24. Mode.Parameter.Statistics.Status
Mode.Parameter.Statistics.MeanN
Mode.Parameter.Statistics.ResTab.Selected
Mode.Parameter.Statistics.ResTab.DelN
Entries for the statistics calculations.
.Status:
On/off switching. Requirement for statistics calculations is a valid
assignment, see 3.2.2.9.
.MeanN:
Number of individual results for statistics calculations.
.ResTab.Select: Selection of the table for the statistics calculations.
original: Original table. The original table is (again) set up, i.e. any
individual results which have been deleted are reincorporated
in the statistics calculations.
delete n: Single result lines are removed from the statistics calcu-
lation. All results of the corresponding line in the statistics
table are deleted. Specification of the line number in
.ResTab.DelN.
delete all: Clear entire statistics table. The results can not be reac-
tivated.
.ResTab.DelN: Specification of the line number to be deleted.
3.2.2.25. Mode.Parameter.Presel.Cond
Mode.Parameter.Presel.DriftDisp
Mode.Parameter.Presel.DCor.Type
Mode.Parameter.Presel.DCor.Value
.Cond: Conditioning ON/OFF
.DriftDisp: Drift display during conditioning ON/OFF.
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of
the drift value at start.
.DCor.Value: Drift value for the manual drift correction.
3.2 Remote control commands
abs., rel., OFF
0...99.99...9999.99
0...±999999
0.01...150, max.
original, delete n,
auto, man., OFF
ON, OFF
2...20
delete all
1...20
ON, OFF
ON, OFF
0.0...99.9
71