RS receive errors:
Parity.
Exit: <QUIT> and ensure settings of appropriate parameters at
both devices are the same.
Framing error.
Exit: <QUIT> and ensure settings of appropriate parameters at
both devices are the same.
Overrun error. At least 1 character could not be read.
Exit: <QUIT>
The internal working-off buffer of the Titrino is full
(>82 characters).
Exit: <QUIT>
RS send errors:
CTS=OFF No proper handshake for more than 1 s.
Exit: <QUIT> Is the receiver switched on and ready to receive?
The transmission of the Titrino has been interrupted with XOFF for
at least 6 s.
Exit: Send XON or <QUIT>
The receive buffer of the Titrino contains an incomplete command
(L
missing). Sending from the Titrino is therefore blocked.
F
Exit: Send L
or <QUIT>.
F
Measuring point list overflow (more than 500 measuring points).
Exit: The error message disappears on next startup.
Missing EP for calculation.
Exit: The error message disappears on next startup or on
recalculation.
No new mean.
Exit: The error message disappears on next startup or on
recalculation.
No new common variable, old value remains.
Exit: The error message disappears on next startup or on
recalculation.
Wrong sample. For KFT with preset titration direction the first
measured value lies behind the endpoint.
Exit: The error message disappears on next startup.
Silo empty and it has been started with open silo or empty silo has
been opened.
Exit: Send a silo entry.
Silo full.
Exit: Send new command.
XXX Bytes are missing so that the method or the silo line could not be stored.
Exit: Send new command.
No new silo result (C24 or C25).
Exit: The error message disappears on next start or on recalculation.
795 KFT Titrino