3.2.2.11. Mode.Parameter.CtrlPara.EP
Mode.Parameter.CtrlPara.UnitEp
Setting of the EP, resp. Control point in mV (with Ipol) or µA (with Upol). The
corresponding unit can be read with .UnitEP.
3.2.2.12. Mode.Parameter.CtrlPara.Dyn
Mode.Parameter.CtrlPara.UnitDyn
Mode.Parameter.CtrlPara.MaxRate
Mode.Parameter.CtrlPara.MinIncr
Control parameters.
.Dyn: Dynamics (control range) in mV (with Ipol) or µA (with Upol). The cor-
responding unit can be read with .UnitDyn.
.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum
possible rate with mounted Exchange Unit.
.MinIncr: Minimum volume increment in µL.
OFF means that the criterion is not monitored.
3.2.2.13. Mode.Parameter.CtrlPara.Stop.Type
Mode.Parameter.CtrlPara.Stop.Drift
Mode.Parameter.CtrlPara.Stop.Time
Mode.Parameter.CtrlPara.Stop.StopT
Type and size of the stop criterion of the titration.
.Type: Type of stop criterion after stop drift or switch-off delay time.
.Drift: Stop drift in µL/min. Applies when "drift" has been selected.
.Time: Switch-off delay time in s. Applies when "time" has been selected.
"inf." means infinite.
.StopT: Stop time in s. Applies when "time" has been selected and the value of
.Time is set to "inf.".
3.2.2.14. Mode.Parameter.TitrPara.Direction
Titration direction.
"auto" means the titration direction is determined automatically by the instru-
ment.
3.2.2.15. Mode.Parameter.TitrPara.XPause
Pause time in s. Runs before dosing the start volume.
3.2 Remote control commands
Ipol:
0...250...±2000
Upol:
0...25...±200.0
Ipol:
1...100...2000
Upol:
0.1...10...200.0
0.01...10...150, max.
0.1...9.9, min.
0...10...999, inf.
0...99999, OFF
read only
read only
drift, time
1...20...999
+, –, auto
0...999999
69