6. CONTINUATION OF THE TECHNICAL MENU

ZERO TRACKING
F5-TRA
Set the zero tracking value.
By default 0,5e.
Possible values from 0.5 to 12.0 in 0.5e increments.
F6-G
GRAVITY
Set the gravity value of the calibration site before calibrating.
Set the gravity value for the destination after the equipment has been calibrated.
F7-BUD
BAUDRATE
Selection of the transmission speed of the RS-232.
Default value 9600bauds
Possible values: 2400, 4800, 9600, 19200
F8-ODE
SENDING MODE
Select the RS-232 send mode.
0
Disabled
1
Manual sending by pressing the key
2
Sending for stability. When the equipment detects stability, it sends the plot 1 time
3
Continuos sending
4
Send on request (without function)
TYPE
F9 -TPE
Select the protocol, at the moment only protocol 1 is available
=
The plot begins with = and is followed by 7 digits including decimal point and negative sign (if any) and in reverse order.
Examples:
Weight 59.723kg
Weight -59.723kg
Weight 59723kg
Same function as point 4 of the manual.
FA-SLP

7. CONNECTIONS

5-pin LOAD CELL connector
n° pin
NAME
1
EXC+
2
EXC-
3
SIG+
4
SIG-
.
W6
W5
Plot =327.950=327.950=327.950=327.950=327.950
Plot =327.95-=327.95-=327.95-=327.95-=327.95-
Plot =3279500=3279500=3279500=3279500=3279500
RS232 4-pin connector
n°pin
1
3
W4
W3
NAME
Tx
GND
OPTIMA USERS GUIDE
W2
W1
EN
EN