12
Valeurs à sécurité intrinsèque
Mode de protection Ex ia
variante de commande
"Sortie"
Option A
Option B
Option C
Option G
Mode de protection Ex ic
Variante de commande
"Sortie"
Option A
Option B
Type de sortie
4-20 mA HART
4-20 mA HART
Sortie impulsion/fréquence/tor
• 4-20 mA HART
• 4-20 mA
PROFIBUS PA
Sortie impulsion/fréquence/tor
Type de sortie
4-20 mA HART
4-20 mA HART
Proline Promass E 200
Valeurs à sécurité intrinsèque
U
= DC30 V
i
I
=300 mA
i
P
=1 W
i
L
= 0 μH
i
C
=5 nF
i
U
= DC30 V
i
I
=300 mA
i
P
=1 W
i
L
= 0 μH
i
C
=5 nF
i
U
= DC30 V
i
I
=300 mA
i
P
=1 W
i
L
= 0 μH
i
C
=6 nF
i
U
= DC30 V
i
I
=300 mA
i
P
=1 W
i
L
= 0 μH
i
C
=30 nF
i
STANDARD
FISCO
U
= 30 V
U
= 17,5 V
i
i
l
= 300 mA
l
= 550 mA
i
i
P
=1,2 W
P
=5,5 W
i
i
L
= 10 µH
L
= 10 µH
i
i
C
=5 nF
C
=5 nF
i
i
U
= 30 V
i
l
= 300 mA
i
P
=1 W
i
L
= 0 µH
i
C
=6 nF
i
Valeurs à sécurité intrinsèque
U
= DC35 V
i
I
= s.o.
i
P
=1 W
i
L
= 0 μH
i
C
=5 nF
i
U
= DC35 V
i
I
= s.o.
i
P
=1 W
i
L
= 0 μH
i
C
=5 nF
i
Endress+Hauser