Handles will not
remain in open
position.
O2 hoop or chest drain
arm will not open or
close.
IVEA will not fully
collapse for storage.
30
IVEA® 600S
Ensure that the blue buttons on each side of the
handle bracket are fully extended and the handles lock
into place with a "click, " before placing weight on the
handles.
To open or close the O2 tank hoop or chest drain arm,
grasp with one hand and depress the corresponding
blue lever. Rotate accessory away from the main column
and up until it "clicks" into its in-use position.
Ensure that the handles, O2 hoop and chest drain arm
are folded in and locked securely against the main
column. Ensure that the IV pole is secured at its lowest
position and the pole adjustment lever is completely
closed. If a power strip is mounted to the IVEA, make
sure cords are wrapped tightly and securely around the
cord hooks and not interfering with the handle locks.