6. Pre-Use or Clinical Use Inspection Points

It is recommended that caregiver performs these inspection points prior to each use.
2X Grips:
in good condition; secure
on handle tubes
IVEA must sit squarely
on floor with no rocking.
Foot Pedal: operates freely (IVEA deployed, no
load); spring returns pedal to raised (locked)
position. Main Column of IVEA should move
downward when forced if and only if Foot
Pedal is depressed
4X Casters: roll & swivel freely; with blue pedal engaged,
both wheel & swivel lock securely; with yellow pedal engaged,
swivel locks in straight forward orientation as shown
If any inspection point fails, contact your hospital biomedical department or
Arjo Technical Support at 1-800-323-1245
12
IVEA® 600S
Elevated IV Pole supports fluid
bags without slipping when
blue lever is closed
Handle Bracket does not slip
downward when blue lever is closed
and weight is applied to handle grips
2X Handle Release Buttons:
pop out fully when handles are
deployed
Orange Lock Pin Indicator:
flush when IVEA deployed
and locked, protrudes
when folded/unlocked