Radiated RF EM fields
Proximity fields
from RF wireless
communications
equipment
RATED power frequency
magnetic fields.
) The equipment is suitable for use in Home Health Environments and Professional Health Care
a
Environments limited to patient rooms and respiratory treatment facilities in hospital or clinics.
The more restrictive IMMUNITY acceptance limits have been considered and applied.
) Before modulation is applied.
b
) This test level assumes a minimum distance of at least 15 cm between the ME EQUIPMENT or ME
c
SYSTEM and sources of power frequency magnetic fields.
Guidance and manufacturer's declaration - Electromagnetic immunity - Input a.c. port
Phenomenon
Electrical fast
transients / Bursts.
Surges Line-to-line
Surges Line-to-
ground
Conducted
disturbances induced
by RF fields.
Voltage dips.
Voltage interruptions
)
c
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
Basic EMC standard
or test method
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
19
)
a
COMPLIANT
NOTE: Further information about distances
to be maintained between portable and
mobile
RF
communications
(transmitters) and the Chicco nebulizers Super
Soft
00009067000000 (Frog) and
00009813000000 (Dog) can be requested from
Artsana S.p.A. using the contact information
provided in this manual. However, it is advisable to
keep the electromechanical aerosol equipment
at an adequate distance of, at least, 0.5 m from
mobile phones or other RF communications
transmitters to minimise possible interference.
30 A/m
)
c
50 Hz or 60 Hz
Immunity test levels
Professional
healthcare facility
environment
± 2 kV
100 kHz repetition frequency
± 0.5 kV, ±1kV
Class II equipment or supplied with Class II AC/
DC adapter. Test not applicable
3V
) 0.15 MHz - 80 MHz
a
6 V
) in ISM bands between 0.15 MHz and 80
a
MHz
)
b
80% AM at 1 kHz
0 % UT; 0.5 cycle
At 0°,45°,90°,135°,180°, 225°, 270° and 315°.
0% UT; 1 cycle and 70% UT; 25/30 cycles
Single phase: at 0°
0% UT; 250/300 cycles.
10 V/m
) 80MHz - 2.7
b
GHz 80% AM at 1kHz
equipment
Home healthcare
environment