TYPE OF DEFECT
Scarce nebulisation
Scarce nebulisation
No nebulisation
No nebulisation
Slow nebulisation
TECHNICAL DATA
Nominal voltage: 230 V~
Service life in storage conditions: 5 years
Absorbed current: 0.6 A
Frequency: 50 Hz
Maximum flow: 14 ± 4 l/min
Fuse: F5 L 250V, 5x20mm
Max pressure: 2.55 ± 0.6 bar
Operational flow: 6 ± 2 l/min
Conditions for using the device: 20' ON - 40' OFF at 40 °C
Operational pressure: 0.76 ± 0.21 bar
MMAD: < 3.7 μm
Weight: 1.6 kg
Nebuliser cup capacity: 10 ml Max; 2ml Min
Breathable fraction: >65%
Appliance service life: 1,000 cycles/treatments
Noise level at 50 cm*: 69.1 dB(A)
Service life of accessories: 1 year
* Data on noise levels were measured on a new device. Values may vary over time with use.
CAUSE
Clogged nebuliser cup
Clogged nebuliser cup
Nebuliser cup components
not assembled correctly
The compressor does not
work
Drug too thick
17
REMEDY
Remove and detach all the
nebuliser cup components
and clean them as
illustrated in this instruction
manual.
If the issue persists after
cleaning, replace the
nebuliser cup.
Disassemble and
reassemble the nebuliser
cup; if the issue persists,
replace the nebuliser cup.
Check that the power cord
plug is disconnected from
the mains and that the on/
off switch is on.
If the issue persists contact
Customer Services.
Dilute the drug as instructed
by the manufacturer.