Chicco SUPER SOFT Mode D'emploi page 16

Masquer les pouces Voir aussi pour SUPER SOFT:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22
CLEANING AND MAINTENANCE OF THE NEBULISER CUP AND ACCESSORIES
• Remove and detach all the nebuliser cup components and dip them in boiling water for about
5 minutes.
• WARNING! Do not boil the masks.
• Rinse the accessories (masks, connector, mouthpiece, fork and tube) under running water. Cold
disinfect all parts that come in contact with the patient using denatured alcohol. Carefully dry
all accessories and the device before reassembling. Store the device and the accessories in a
cool, dry place, away from light and heat.
• Never use benzene, thinners or other inflammable chemical substances for cleaning.
• For increased hygienic safety, never use the same accessories on more than one patient. Pur-
chase a dedicated kit for each user (Nebuliser Kit
• Clean and check that the condition of the accessories before and after use. Replace them if
damaged.
CHECKING AND REPLACING FILTERS
Periodically check the condition of the filter. The filter was inserted to protect the compres-
sor. Correct filter maintenance prolongs the life of the aerosol. The frequency with which filter
should be replaced depends on the conditions in which the device is used. The filter is located
on the back of the device.
TO REPLACE THE FILTER (fig. 2):
• Remove the filter cover cap turning it anti-clockwise;
• Remove the filter to be replaced from its housing (using a small tool if necessary);
• Carefully fit the new filter in its housing;
• Replace the filter cover cap and fasten tightly by turning it in a clockwise direction;
• WARNING! Never leave the removed filter cover cap and/or filters unattended: These
are small parts posing the risk of suffocation if swallowed by children.
00009069000000).
Super Soft
16

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

0000906700000000009813000000

Table des Matières