Description
Char
Peak hold time
P
1
Power / AD value
Q
Presel. Setpoint
SV
Target temperature
T
Get Analog Input 1
TV1I
Get Analog Input 2
TV2I
Temperature Unit
U
Poll / Burst mode
V
Command counter
W
Burst string contents
X$
Multidrop Address
XA
Profibus / Modbus
XAS
Address
Device Bottom range
XB
limit
Restore Box Factory
XF
defaults
Transmission
XG
Device High range limit XH
Sensor initialization
XI
Box Temperature
XJ
FTC3 trigger/hold
XN
analog Output 1 mode
XO1O
(XO)
setting average / peak / valley / advanced hold cancels all other hold modes
1
for MI3COMM only
2
Mode 0 and 4 for MI3MCOMMA only
3
Mode 5 to 8 for MI3COMM only
4
MI3
Format
P B S N Legal Values
nP
0.0 - 998.9s
float
?nQ
nSV
float
?nT
float
?TV1I
?TV2I
X
C / F
V=P
?W
?X$
0nn
n
?nXB
float
float
?nXH
float
float
nXN
XO1O=v
(XO=v)
2
Rev. G3 Oct 2020
ASCII programmation
(999 = )
answer !nT=0099.9
0 – 5V
0 – 5V
P=poll B=burst
1 – 0x7FFF (repeat)
Set to 1 when V=P
000 – 032
(000 --> single unit mode)
0 – 125 (Profibus)
1 – 247(Modbus)
answer !nXB=10.0
0.1 – 1.0
read only
1 after reset
0 if XI=0
(C/F)
T = trigger, H = hold
0 – 0...20 mA
3
4 – 4...20 mA
5 – TCJ
4
6 – TCK
7– TCR
8 – TCS
9 – 0...5 V
10 – 0...10 V
99 – disable (tristate)
Factory default Head /
Box
0.0s
H
H
500°C
H
H
B
B
C
B/H
poll mode
B
1
B
B
000
B
0 (Profibus)
B
1(Modbus)
H
except address B
1.0
H
H
1
B
B
T
H
XO1O=9 for
B
communication
box (metal)
XO1O=99 for
communication
box 6TE, analog
129