RS232 Interface
Output examples (space indicated by '*'):
********0.00*g*****
*******12.73*g****?
*******12.72*g****?
******0.4500*oz****
******0.5950*oz***?
***********3*PCS**?
***********3*PCS***
********0.85*oz*****WET*WT
********0.85*oz*****AIR*WT
Output Format 2 (for the models 401/601/6000):
Output String:
[weight]
[space]
[unit]
[space]
[stability indicator]
[space]
[Legend]
Note: The unit field length varies with different units. The weight field could be 11 or 12, depending
on if the weight string has an additional decimal point.
Output example (space indicated by '*'):
********200*g**
********311*g*?
*******0.211*kg*?
Navigator (NV/NVL/NVT) Output:
Output String:
[weight]
[space]
[unit]
[space]
[stability indicator]
[space]
[NET]
[Legend]
Note: The unit field length varies with different units, the Legend field is only printed in certain modes,
and the time string only printed at internal print.
11~12 characters (right justified)
1 character
1~5 characters
1 character
1 character; "?" when unstable, blank when stable
1 character
10 character
10 characters (right justified)
1 character
1~5 characters (left justified)
1 character
1 character; "?" when unstable, blank when stable
1 character
0 or 3 character
0~16 character
EN-3