SPAN
2. When
is displayed, press
the YES button to access the Span
Calibration menu item. The display
0
flashes
.
3. With no weight on the scale, press
the YES button to establish the zero
point. The display shows
the zero point is established.
4. The display flashes the span
calibration point. Place the
specified weight on the scale and
press the YES button.
5. To choose a different span point,
repeatedly press the NO button to
increment the selections or press the BACK button to
decrement the selections. When the desired value is
displayed, place the specified weight on the scale
and press the YES button.
--C--
6. The display shows
the span point is established.
7.
If span calibration was successful,
the scale exits to the active
weighing mode and displays the
actual weight value.
LINEARITY CALIBRATION
1.
Linearity calibration uses three calibration points.
The first calibration point is established with no
weight on the scale. The second calibration point
is established at approximately half capacity. The
third calibration point is established at capacity. The
linearity calibration points are fixed and cannot be
altered by the user during the calibration procedure.
L INEAr
2. When
is displayed, press
the YES button to access the
linearity calibration menu item.
3. The display flashes
weight on the scale, press the YES
button to establish the zero point.
--C--
4. The display shows
zero point is established.
5. The display flashes the mid
calibration point.
6. Place the specified weight on the
scale and press the YES button.
--C--
The display shows
mid point is established.
PAGE 4 OF 24
OPERATION CONTINUED
--C--
while
while
0
. With no
while the
while the
7. The display flashes the
full calibration point.
8. Place the specified weight on the
scale and press the YES button.
--C--
The display shows
full point is established.
9. If linearity calibration was
successful, the scale exits to
the active weighing mode and
displays the actual weight value.
ZERO OPERATION
Zero can be set under the following conditions:
• Automatically at power on (initial zero).
• Semi-automatically (manually) by pressing the
ON/ZERO OFF button.
• Semi-automatically by sending the zero command
(Z or alternate zero command).
Press the ON/ZERO OFF button
to zero the weight display.
The scale must be stable to
accept zero operation.
MANUAL TARE
1.
When weighing an item
that must be held in a
container, taring stores the
container weight in memory.
Place empty container on the scale (example 0.5 kg.)
and press the TARE button. Display will show net
weight.
2. To clear the tare value,
empty the scale and press
the TARE button. The display
will show gross weight.
CHANGING UNITS OF MEASURE
Press and hold the PRINT UNITS button until the desired
measuring unit appears.
PRINTING DATA
1.
Printing the displayed data to a printer or sending
data to a computer requires the communication
parameters are set in the print menu.
while the
0521 IH-2587