Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to
maintain a device in accordance with the instructions may invalidate the device's warranty. Furthermore,
failure to maintain a device may compromise the clinical condition or safety of users and/or their
caregivers.
1st Day
EN
• Read the User Guide carefully before
use and store it in a safe place for future
reference
Before every use
• Check that all castors are moving freely
and that all wheel locks/brakes are easily
activated
• Make sure all cables are securely plugged in
before use in the shower
• Inspect all cables for cracks or signs for
worn-out parts. Never use a product which
appears to be substandard or faulty. Contact
your local dealer for service
• Disinfect the upholstery. please check the
disinfection section in this User Guide for
more information
• Inspect the detachable parts. If required,
disinfect the parts with a 70% disinfectant
solution. It is recommended to wipe of any
residue and dirt using a cloth wrung in warm
water and a mild detergent/soap without
chlorine. Let the detachable parts dry before
disinfection
After every use
• Wipe the product dry
Daily
• Visual inspection. Repair or replace any
damaged or worn-out parts
• Ensure that all fasteners are in place and
fitted correctly
• Check all fittings and buckles for signs of
worn out parts
• Be aware of any MAX mark indications
before adjustment
Monthly
• Use a cloth with warm water and a mild
detergent to clean the product
• All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
• Oil the swing-away parts. We recommend
you use a professional lubricant system
Annually
• Inspect the frame for cracks or signs for
worn-out parts and carry out annual service.
Never use a product which appears to be
substandard or faulty
• Inspect the piston rods are greased, extract
the piston rods to check grease
Warewashing machine
This product can be washed at:
• 60°C in 10 minutes
• 85°C in 3 minutes
...using a mild detergent in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product. Clean the
accessories separately.
Before washing
• Avoid degreasing of the piston rods, the
actuators should be retracted to minimum
stroke and without load before washing.
• Remove all accessories, unplug all cables
and make sure to dismount the battery,
control-box and handset. Leave the
actuators with the cables mounted on the
product (see page 22)
After washing
• Ensure correct placement of cables inserted
into the control box when reinstalling the
power unit system (see page 23)
• Check that the handset functionalities works
correct
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to
wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before disinfection.
Service interval
A detailed inspection of the product must be
made every 12 months (every 6 months for heavy
users) and every time the product is reissued for
use. The inspection should be carried out by a
technically competent person who has been
trained in the use of the product.
The lifetime of this product in normal use is 5
years if all maintenance and servicing is carried
out in accordance with the manufacturer's
instructions and demonstrably recorded.
34
support.R82.org