Operation cont.

Fuse Replacement

WARNINGS:
To avoid electric shock, remove the test leads from
the meter before removing the battery/fuse cover.
1. Lift up the tilt stand.
2. Loosen the Phillips
screws on the battery/fuse
cover.
3. Remove the battery/fuse
cover.
4. Gently remove fuse and
install new fuse into the
holder.
5. Always use a UL recognized
fuse of the proper size and
value: 10A/600V (5 x 20mm)
fast blow.
6. Install the battery cover and
tighten the screws
WARNING:
To avoid electric shock, do not operate meter until
the battery/fuse cover is securely fastened to the meter.
WARNING:
When replacing the battery or fuses, be sure to secure
the battery compartment door firmly to maintain the waterproof and dust
proof integrity of the meter. Loose or overtightened screws, or an improperly
seated o-ring may compromise the meter's water and dust ingress protection.
17
Specifications
Accuracy is given at 65°F to 83°F (18°C to 28°C), less than 70% relative humidity
Function
AC Voltage
All AC voltage ranges are specified from 5% of range to 100% of range
Input Protection:
Input Impedance:
AC Response:
Function
Low Z
AC Voltage
All AC voltage ranges are specified from 5% of range to 100% of range
Input Protection:
Input Impedance:
AC Response:
Function
DC Voltage
Fuse
Input Protection:
Input Impedance:
Function
Low Z
DC Voltage
Input Protection:
Input Impedance:
Range
Resolution
Accuracy ±
±(1.0% +5 digits)
4.000V
1mV
40.00V
10mV
±(1.2% +5 digits)
400.0V
0.1V
±(1.5% +5 digits)
600V
1V
600V AC RMS or 600V DC
10M
50 to 60Hz
Range
Resolution
Accuracy ±
4.000V
1mV
40.00V
10mV
400.0V
0.1V
300V AC RMS or 300V DC
approx. 3k
50 to 60Hz
Range
Resolution
Accuracy ±
400.0mV
0.1mV
4.000V
1mV
40.00V
10mV
400.0V
0.1V
600V
1V
600V RMS or 600V DC
10M
Range
Resolution
Accuracy ±
400.0mV
0.1mV
4.000V
1mV
40.00V
10mV
400.0V
0.1V
300V AC RMS or 300V DC
approx. 3k
18
(% of reading + digits)
(% of reading + digits)
±(1.5% +5 digits)
(% of reading + digits)
±(1.0% + 8 digits)
±(1.0% + 3 digits)
±(1.2% + 3 digits)
(% of reading + digits)
±(1.5% + 5 digits)