C a l v a l u e s a r e
s e t t o d e f a u l t
Cause
The RAM and EEPROM contain
different calibration data, since
calibration data was not saved after
a change.
Solution
The calibration data is transferred
automatically from the EEPROM to
the RAM, and thus the old
calibration values are restored.
All manuals and user guides at all-guides.com
O n l i n e M o d e
K e y b o a r d i n a c t i v e
Cause
The instrument is in Online Mode,
and is therefore being controlled
from the PC. Only the <operate>
and <reference> buttons and the
foot switch are active.