MP85A(DP) FASTpress, MP85A(DP)−S EASYswitch
In this example, a displacement transducer is connected to channel x and a
force transducer to channel y.
Step 1 (transducer window):
The displacement transducer works as a half bridge with an 80 mV/V mea-
suring range (the next highest with 100 mV/V was selected).
The force transducer works as a full bridge with an 2 mV/V measuring range
(the next highest with 4 mV/V was selected).
It is helpful to set the units at the same time (here: mm and kN).
Step 2 (enter characteristic curve window):
Two methods are available:
− Enter characteristic curve using the data sheet data
− Alternatively: Calibrate characteristic curve, specifying two points per sen-
sor