Télécharger Imprimer la page

Steinbach SBS-LZ-1000SLS Manuel D'utilisation page 24

Centrifugeuse de laboratoire

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 51
immediately accessible from the area where the device is being used. Make sure
that the pressure of the air supplying the device does not exceed the
recommended range!
ASSEMBLING THE DEVICE
Place the device on a flat, stable surface so that all four rubber feet are firmly set
on the surface.
Connect the power cord to the device and its other end to a 230VAC socket. Please
note: The power socket must be grounded. This will reduce the risk of electric
shock.
3.3.
1. Check that the nut on the rotor has not loosened, if so, tighten it.
2. Get the test tubes with a preparation ready for centrifugation and place them
in the device so that they meet the following premises:
- the test tube should be filled to a maximum of 75% of its total volume,
- place the test tubes symmetrically in the rotor.
- The sample-filled tubes placed opposite each other must be of equal weight.
- Always verify the condition of the test tube used. Do not insert cracked or deformed
test tubes in the device!
3. Close the device lid.
4. Switch the device connected to the power supply on, using the ON/OFF button at
the back of the device.
5. Set the device parameters.
Press and hold the (4) button on the control panel until "A1 ※ 3100" flashes on the
display. Then, use the "Up" (5)/"Down" (6) buttons to set the preferred speed value.
The maximum speed must not be exceeded.
Device use
Speed setting

Publicité

loading