Technical Specifications; Main Performance - PULOX PO-650B Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

b. Relative Humidity :≤75%
c. Atmospheric pressure:700hPa~1060hPa
3. Principle
Principle of the Oximeter is as follows: An experience formula of data
process is established taking use of Lambert Beer Law according to
Spectrum Absorption Characteristics of Reductive Hemoglobin (Hb)
and Oxyhemoglobin (HbO2) in glow & near-infrared zones. Operation
principle of the device is: Photoelectric Oxyhemoglobin Inspection
Technology is adopted in accordance with Capacity Pulse Scanning &
Recording Technology, so that two beams of different wavelength of
lights can be focused onto human nail tip through perspective clamp
finger-type sensor. Then measured signal can be obtained by a
photosensitive element, information acquired through which will be
shown on screen through treatment in electronic circuits and
microprocessor.

4. Technical Specifications

4.1 Main Performance

A. SpO2 value display
B. PR value and bar graph display
C. Pulse waveform display
D. Low-battery indication: low-battery indication appears before
working abnormity which is due to low-power
E. Screen brightness can be adjusted
F. PR sound indication
G. With alarm function
H. With SpO2 value and pulse rate value record function,the stored
data can be uploaded to computer
I. It can be connected with an external oximeter probe
J. Real-time data uploading function
K. Review function
Figure 1