5.1 Cleaning/disinfection
Note
Always follow disposal and hygiene instructions and
regulations!
Cleaning/disinfection
Risk of electric shock and fire haz-
ard!
Unplug the device before cleaning.
DANGER
Ensure that no fluids will pene-
trate the product.
Leave the product exposed to
air for at least 1 minute after
cleaning/disinfecting.
Damage to the product due to
incorrect cleaning or inappropriate
cleaning/disinfecting agents!
CAUTION
Use any commercial detergent
for surface cleaning.
Never immerse the product in
water or detergent.
Do not introduce any fluid into
the charging bays.
Clean the contacts in the charg-
ing bays, applying extreme care.
Wipe the shell of the product with a lint-free cloth
moistened with a commercial disinfectant for wipe
disinfection.
If necessary, after the appropriate exposure time,
wipe off any residues of the cleaning/disinfecting
agent with a lint-free cloth moistened with clean
water.
Dry the product with a clean, lint-free cloth.
Clean the contacts in the charging bay with isopro-
panol or ethyl alcohol and a cotton swab. Do not
use corrosive chemicals.
If necessary, repeat the cleaning/disinfecting pro-
cess.
All manuals and user guides at all-guides.com
6.
To ensure reliable operation, Aesculap recommends
servicing the product at least once a year.
Service addresses
In the UK
KG PRODUCTS
243-251 City Road, Fenton
Stoke-on-Trent, Staffs
ST4 2PX
United Kingdom
Phone:
DIAMOND EDGE
126 Gloucester Rd.
Brighton, BN1 4BU
United Kingdom
Phone:
Or in the US
A & J Sales and Service
42 - A Frame Drive
West Burke, VT 05871
USA
Phone:
Or in Germany
Albert Kerbl GmbH
Felizenzell 9
84428 Buchbach / Germany
Phone:
Fax:
E-Mail:
www.kerbl.com
Other service addresses can be obtained from the
address indicated above.
Maintenance
(44) - 01782 844 866
(44) - 01273 605 922
(800) 447-1957
+49 (0) 8086 933-100
+49 (0) 8086 933-500
info@kerbl.com
7