Over the frequency range 150 KHz to 80 MHz, field strengths should be
less than 3V/m.
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM
for EQUIPMENT or SYSTEM that not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the CMS50D
Pulse Oximeter
The CMS50D Pulse Oximeter is intended for use in the electromagnetic
environment in which radiated RF disturbances are controlled. The
customer or the user of the CMS50D Pulse Oximeter can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
CMS50D Pulse Oximeter as recommended below, according to the maximum
output power of the communications equipment.
Rated
maximum
output power
of transmitter
(W)
0.01
0.1
1
10
100
Separation distance according to frequency of
transmitter
(m)
150KHz to
80MHz
3
5 .
d
P
E
1
0.12
0.37
1.17
3.69
11.67
80MHz to
800MHz
3
5 .
d
P
E
1
0.12
0.37
1.17
3.69
11.67
800MHz to
2.5GHz
7
d
P
E
1
0.23
0.74
2.33
7.38
23.33