Air Filter - FLAEM RespirAir Mode D'emploi

Appareil pour aérosolthérapie
Masquer les pouces Voir aussi pour RespirAir:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 11
CLEANING IN A CLINICAL OR HOSPITAL SETTING - DISINFECTION AND STERILISATION
Before disinfection or sterilisation, sanitise the nebuliser cup and the accessories, choosing
one of the methods described below.
method A: Sanitise accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4 under potable hot water
(approximately 40°C) with a gentle, non abrasive dish detergent.
method B: Sanitise accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4 in the dishwasher with a
hot cycle.
DISINFECTION
Accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4 can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
If you want to also perform the STERILISATION, jump to the STERILISATION paragraph.
After having disinfected the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
STERILISATION
Accessories C1 (Assembled nebulizer)-C2-C3-C4 can be sterilised.
Device: Fractionated vacuum overpressure steam steriliser in accordance with EN 13060.
Implementation: Wrap every single part to be treated with a sterile barrier system or
packaging in accordance with Norm EN 11607. Place the packed components in the steam
steriliser. Run the sterilisation cycle according to the operating instructions of the device by
selecting a temperature of 134°C and a time of 10 minutes first.
Storage: Store the sterilised parts as per the instructions for use of either the sterile barrier
system or packaging.
The sterilisation procedure is validated in its conformity to ISO 17665-1.
At the end of each use store the device complete with accessories in a dry place away
from dust.

AIR FILTER

The device has a suction filter (A3) that should be replaced when it is
dirty or changes color. Do not wash or reuse the same filter. Regular
replacement of the filter is required to help and ensure correct
performance of the compressor. Check the filter on a regular basis.
Do not replace the filter during use.
Contact your local dealer or authorized service center for spare
filters. Only use original accessories and spare parts by Flaem, we disclaim any
liability in the event of using non original spare parts or accessories.
5

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

P0611em f1000Respir airHospineb

Table des Matières