Internal temperature change greater than a pre-set amount (typically 2ºC for
Precision balances).
Time since last calibration exceeds a pre-set time (typically 4 hours, or 15 minutes
after power is applied).
The balance will indicate the need for calibration to be carried out by flashing the "
symbol on the display. As soon as the balance is calibrated the symbol will be turned off.
The Auto calibration feature can be enabled, disabled or changed within the user options
to meet the requirements of the users.
6.2.5 Calibration Errors
Occasionally during calibration an error will be detected. These errors can be caused by:
Unstable readings
Improper calibration weights being used
Large shifts of zero from the factory settings
When an error is found a displayed message will be shown and the calibration must be
done again. If the balance has error messages more than once it is possible the
mechanics have been damaged.