1.2. TROUBLESHOOTING

Mistake
Brake function does not work
Upper part hangs skewed to one side
Upper part is stuck
The gait trainer rolls badly
The handle is loose
The seat height and gait height are lower than before
The seat height is lower than before
If the above does not solve your problem, you should contact your VELA
supplier.
Possible cause and solution
The brake is not activated. See section 2.0.3.
Check that both springs are fitted correctly. See section
2.0.2.
Check that the upper part is set correctly and is the
same height on both risers. See height marking in the
inner tube. See section 2.0.2.
Wheels are blocked. Clean the wheels for wire residue
and the like.
Tighten the screws
Loosen the top Allen screws and hand screws on the
upper part and move the upper part to the desired
position and tighten the hand screws and Allen screws
again. See section 2.0.5./2.0.6.
The tilt bar is folded down. See section 2.0.10.
E N