Marks and symbols
Symbol
Description
Network connection: LAN interface to the PC
Interface to the
prismaCONNECT module
Do not dispose of the product in domestic
waste
Degree of protection against electric shock:
Protection class II product
Protect from moisture
Permitted temperature range for transport
and storage
Permitted humidity range for transport and
storage
Manufacturer and date of manufacture
Follow the instructions for use
Serial number
Order number
Indicates the product is a medical device
Unique device identifier
CE symbol (confirms that the product con-
forms to the applicable European directives/
regulations)
5
Operation
5.1 Connecting the module
1. Make plug connections in line with Figure 1 or
Figure 2.
You can use the module with the following
devices:
•
prisma SMART series (WM090TD)
•
prismaLINE series (WM100TD) (in
conjunction with the prismaCONNECT
module)
•
prisma VENT series (WM110TD / WM120TD)
(in conjunction with the prismaCONNECT
module)
Do not connect several devices simultaneously.
The relevant LED comes on.
2. Attach ferrites to the cables in line with Figure 1
or Figure 2, attaching the ferrites as close as
possible to the prisma HUB module.
5.2 Switching on the module
1. To switch on the module: connect the module to
a therapy device.
The LEDs which have a plug connection come on.
The LEDs go out one minute after the last
change.
2. To switch on the LEDs: press PSG key 8..
5.3 Outputting a PSG test signal
1. Press PSG key 8..
The LED comes on. A synthetic signal is output to
the PSG system for 2 minutes.
5.4 Configuring the module
The DHCP preset is activated at the LAN port for con-
nection to a PC 2.. A dynamic IP address is assigned
to the prisma HUB module. If no DHCP server is avail-
able, the IP address is assigned by zeroconf/auto-IP in
the range 169.254.1.0 to 169.254.254.255.
You can assign a static IP address to the module with
the aid of the prisma HUBconfig PC software.
5.5 Integrating the module in an IT
network
The module can be integrated in an IT network to cre-
ate a link between the module and a PC. The network
must have the following properties:
•
UDP and TCP ports 51337 and 51338 must be
open and may not be blocked by a firewall.
•
The UDP protocol is required to identify devices in
the network automatically, data exchange
subsequently takes place using the TCP protocol.
If the IT network fails, you can continue operating the
therapy device via the display and transmitting data
via SD card. This prevents failure of the IT network
giving rise to hazardous situations.
6
Hygiene treatment
6.1 General information
•
Wear appropriate safety gear for the disinfecting
process.
•
Refer to the Instructions for Use for the
disinfectant used.
6.2 Intervals
Clean the module weekly. If required (e.g. following
infectious diseases or severe contamination), disinfect
the module. Sterilizing is not permitted.
6.3 Subjecting module to a hygiene
treatment
Risk of material damage as a result of ingress of
liquids!
The module may be damaged by the ingress of
liquids.
Do not immerse the module in liquids.
1. Detach all connections from the module.
2. Wipe the module over with a soft cloth.
3. If necessary: remove more stubborn particles of
dirt with a soft cloth moistened with water or
disinfectant (recommended products: terralin
protect or perform advanced Alcohol EP).
7
Function check
Carry out a function check after every hygiene treat-
ment and repair, but at least every 6 months.
1. Remove all cables from the module one after
another.
The relevant LED goes off.
2. Re-connect all the cables to the module one after
another.
The relevant LED comes on.
3. Press PSG key 8..
The LED comes on. A synthetic signal is output to
the PSG system for 2 minutes.
4. If the module is not working properly or is
damaged: contact your specialist dealer.
8
Troubleshooting
Fault
Cause
Remedy
Error LED 7. is
Technical fault
Have module repaired.
on
No LAN
Check whether the LAN
Cable plugged in
connection
cable is in the socket
incorrectly
possible
marked
No connection
Check whether the PSG
to the
Cable plugged in
cable and the LAN ca-
prisma-
incorrectly or not at
ble are in the sockets
CONNECT
all.
marked
module
LEDs are on.
Check whether the PSG
No connection
Cable plugged in
cable is in the socket
to the prisma-
incorrectly
marked prismaPSG
PSG module
and the LED is on.
IP address of the
Check IP settings with
prismaHUB config-
the aid of the prismaTS
ured incorrectly
software.
No therapy
The prisma-
device is
Configure prisma-
CONNECT module
found in the
CONNECT module to
was configured to a
prismaTS
DHCP
static IP address
software
Enable UDP and TCP
Firewall blocking
ports 51337 and
data transmission
51338.
9
Servicing
The module is designed for a service life of 6 years. If
used in accordance with its intended use, the module
requires no maintenance according to these Instruc-
tions for Use.
®
10 Scope of supply
Part
Item number
prisma HUB, basic device
WM 31665
HUB connecting cable
WM 31664
PSG module connecting cable
WM 29696
3 m CAT-6 patch cable, green
WM 98273
Ferrite (3x)
WM 31669
Key for ferrite
WM 31659
Instructions for Use
WM 68240