Technical description
5
Technical description
5.1
Function description
The device is a refrigerator with absorption cooling. Ammonia is used as the refriger-
ant.
All materials used in the refrigerator are compatible for use with medicines and
medicinal products.
The refrigerant circuit is maintenance-free.
The refrigerator is intended for freestanding use. The cooling unit cover protects the
warm parts of the cooling units. The foot frame with ventilation openings ensure
sufficient ventilation.
The electronics checks the refrigerator's function. If it registers a problem it shuts
down the refrigerator and the interior light flashes continuously.
Two door sensors check whether the door is open. If the door is kept open for more
then 20 s, the refrigerator emits an audible warning signal. After another 40 s it emits
another audible warning signal.
The separate digital thermometer shows and stores the minimum and maximum
inner temperatures from the last reset so that you can check whether the medicines
have been stored at the appropriate temperatures.
5.2
Control elements
No. in
fig. 1, page 3,
fig. 2, page 4
1
2
3
4
5
6
7
8
14
Explanation
Digital thermometer
Interior light
Cover for cooling unit
Evaporator
Door
Door lock (optional)
Air sensor
Foot frame
DS301H, DS601H
EN