[Compliance chart]
Diameter of balloon inflated by
recommended inflation pressure (NP)
Balloon Inflation Pressure
Balloon
Diameter
MPa
0.2
0.4
(mm)
2
4
atm
1.00
0.97
0.98
1.00
1.20
1.18
1.20
1.20
1.50
1.44
1.47
1.50
1.75
1.65
1.70
1.75
2.00
1.86
1.93
2.00
2.25
2.08
2.16
2.25
2.50
2.29
2.39
2.50
2.75
2.53
2.64
2.75
3.00
2.77
2.88
3.00
3.25
2.99
3.12
3.25
3.50
3.20
3.35
3.50
3.75
3.44
3.59
3.75
4.00
3.67
3.82
4.00
This data is reference value, not guaranteed value.
Diameter of balloon inflated by
maximum inflation pressure (RBP)
NP
RBP
0.6
0.8
1.0
1.2
1.4
6
8
10
12
14
1.02
1.05
1.07
1.09
1.22
1.24
1.26
1.28
1.53
1.55
1.58
1.61
1.79
1.82
1.85
1.89
2.05
2.09
2.12
2.16
2.31
2.35
2.40
2.44
2.57
2.62
2.67
2.72
2.83
2.88
2.93
2.99
3.09
3.15
3.20
3.25
3.34
3.41
3.48
3.53
3.60
3.68
3.75
3.81
3.88
3.97
4.04
4.11
4.15
4.26
4.34
4.41