Télécharger Imprimer la page

Masimo RD SET SL Serie Mode D'emploi page 3

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 6
RD SET™ SL Series
Patient Cables
Reusable
Prior to using this Cable, the user should read and understand the Operator's Manual for the Device, this Directions
for Use and the Directions for use for the Sensor.
INDICATIONS
The RD SET™ SL Series patient cables have the same indications for use as the applicable sensors. Please see the compatible
sensor's directions for use for indications for use and prescribed information.
DESCRIPTION
The RD SET SL Series patient cables are used with Masimo® RD SET sensors.
The RD SET SL Series patient cables and Masimo sensors have been functionally verified on Spacelabs devices with Masimo
SET® technology.
RD SET SL Series patient cables have been verified using Masimo SET technology.
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET oximetry or licensed to use
Masimo sensors.
WARNINGS, CAUTIONS, AND NOTES
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and
discard if cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and
sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable may result in intermittent readings,
inaccurate results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting
either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize the
cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage
the electrical components, potentially leading to patient harm.
• Caution: Replace the cable when a replace cable message is displayed, or when a low SIQ message is consistently
displayed while monitoring consecutive patients after completing the low SIQ troubleshooting steps identified in the
monitoring device operator's manual.
• Note: The cable is provided with X-Cal™ technology to minimize the risk of inaccurate readings and unanticipated loss
of patient monitoring. The cable will provide up to 17,520 hours of patient monitoring time. Replace the cable when the
patient monitoring time is exhausted. Some legacy monitoring devices do not distinguish between the cable and the
sensor. In these devices, a replace sensor or similar message could pertain either to the sensor or the cable.
INSTRUCTIONS
A) Attach the RD SET SL Series patient cable to the Device
1. Orient the connector of the cable to mate with the patient cable connector on the Masimo SET device and insert.
B) Attaching the RD SET SL Series patient cable connector to an RD SET disposable sensor connector
1. Refer to Fig. 1. Orient the sensor connector to the patient cable connector as shown.
2. Refer to Fig. 2. Insert the sensor connector completely into the patient cable connector until it locks in place as shown.
C) Attaching the RD SET SL Series patient cable connector to the RD SET reusable sensor connector
1. Refer to Fig. 3. Orient the sensor connector to the patient cable connector, as shown.
2. Refer to Fig. 4. Insert the sensor connector completely into the patient cable connector. Close the sensor connector
cover over the patient cable connector until it locks in place.
D) Disconnecting the RD SET SL Series patient cable connector from the RD SET disposable sensor connector
1. Refer to Fig. 5. Pull firmly on the sensor connector to remove it from the patient cable.
DIRECTIONS FOR USE
LATEX
Not made with natural rubber latex
PCX-2108A
02/13
3
en
Non-Sterile
9276D-eIFU-0118

Publicité

loading