8.3 Measurement

Warning:
For reliable measurements the relative humidity must not exceed 60%.
Note
− To obtain proper measurements, earthing of the measuring instrument is essential.
− The buzzer sounds continuously, when a value higher than 330 volts is displayed
(overrange). Stop measurement in this case.
1 Earth the ion balance plate to remove any remaining static charge.
2 Earth the conductive case with the supplied grounding lead.
3 Switch the instrument on to measure the ion balance: press the [POWER] button once and
then press the [MODE] button until the display shows "Ib". The distance indicating LED's do
not light up.
4 Point the measuring instrument to an earthed surface.
5 If reading > zero: press the [ZERO] button.
6 Hold the measuring instrument in the ionized air flow to be measured.
7 The measuring instrument indicates the voltage measured (V) and the polarity.

9 Maintenance

9.1 General

− Keep the measuring instrument clean.
− Replace the battery when indicated by the battery indicator [5.3].

9.2 Replacing the battery

Warning:
− Do not pull the wires of the battery clip.
− Pay attention to polarity when connecting the battery.
− Do not exert pressure on the battery cover. This may damage
the cover.
− Make sure no wires are pinched between the casing and the
battery cover when closing it.
1 Remove cover at the rear.
2 Remove the old battery.
3 Connect the new battery and insert it into the battery compartment.
4 Close cover.
Hint:
− If the measuring instrument will not be used for a long period: remove the battery.