Figure 1, FMX-004
This fieldmeter can also be used for ion balance voltage measurements. The ion balance
voltage can be measured up to +/-300 V by connecting the ion balance pllate to measure the ion
balance and then pressing the light blue [MODE] but on. The plate can be kept attached to
either end of FMX-004 when not in use. The FMX-004 is switched off automatically
approximately 5 minutes after it has been switched on. This feature can be switched off, simply
press and hold the red [POWER] button for at least three seconds when turning the FMX-004
on. Three led bars, the A. OFF sign and three beeps indicate that the feature automatic switch-
off has been disabled.
Dahecinst
L ED indicators
Sensor
Ib-adapter
contacto@dahecinst.com
dahecinstrumentacion@gmail.co
Buttons (4x)
+52 5586114372
+52 7791340470
Release button
Ib-adapter
Earth connection
and analog out
Release button
Ib-adapter
GB