The oven must be installed in a protected area using guards equipped with systems
that allow only frontal access (gates, shutters etc.).
Position the equipment so that the fan flue is beneath the mobile hood
aspiration tube.
The aspiration fan flue must be free and must not be connected to other tubes or hood ducts.
Should the above-described connection be necessary it is essential that no elbows or
bends be created as combustion fumes and gases would no longer be drawn
outwards.
6.3.1 Worktops
156 of 272
MAINTENANCE AND USE MANUAL
EN
The equipment (the example in the
figure
is
the
DM
designed for use on a rigid worktop parallel
to the floor. Suitable worktops include
service
furniture
operator is standing) or modelling tables
(where work is done sitting down). All
worktops must be checked for stability.
40)
has
been
(where
the
FURNACE DM