Installation
3.2.2 Panel mounting, installation dimensions
• Installation depth: approx. 171 mm (6.73 inch) (incl. terminals and fastening clips)
• Panel cutout: 138
• Panel thickness: 2 to 40 mm (0.08 to 1.57 inch)
• Maximum point of view range: From the display center 50° to the left and right, 20° upward,
30° downward.
• Securing to DIN 43 834
1.
Push the device through the panel cutout from the front. To avoid heat accumulation, we
recommend keeping a distance of > 15 mm (> 0.59 in) from the walls and other devices.
2.
Hold the device level and then hang jack screws in the openings (2 x top, 2 x bottom).
3.
Tighten the screws on the jack screws equally with a screwdriver, such that a secure seal
to the control panel is guaranteed.
!
Note!
A distance of min. 15 mm (> 0.59 in) between the devices has to be observed if aligning the
devices in the Y-direction (see Fig. 2, vertically above one another).
The devices can be arranged horizontally beside one another in the X direction (see Fig. 2) with-
out any spacing between the devices.
3.3
Mechanical locking
The CompactFlash card can be secured
against unauthorized removal using the
locking plate supplied. For this purpose,
insert the locking plate into the slit below the
handle of the cover plate (see Fig. 3). Now
you can lead-seal or seal the cover plate with
a suitable lock.
26
+1
+1
x 138
mm (5.43
+0.04
+0.04
x 5.43
inch)
Fig. 3: Inserting the locking plate
Multi Channel Recorder