Technical specifications
Compatible devices
For a full list of compatible devices for this mask, see the Mask/Device Compatibility List
at ResMed.com/downloads/masks. SmartStop may not operate effectively when using this
mask with some CPAP or bi-level devices.
Pressure-flow curve:
Mask setting options:
For AirSense, AirCurve or S9
Therapy Pressure:
Resistance: Drop in pressure measured (nominal)
at 50 L/min
at 100 L/min
Sound: Declared dual-number noise emission values in accordance with ISO4871:1996
and ISO3744:2010. A-weighted sound power level and A-weighted sound pressure level at
a distance of 1 m, with uncertainty of 3 dBA, are shown
Sound power level (QuietAir vent)
Sound pressure level (QuietAir vent)
Sound power level (Standard vent)
Sound pressure level (Standard vent)
Environmental conditions:
Operating temperature: 5°C to 40°C
Operating humidity: 15% to 95% non-condensing
Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: up to 95% non-condensing
Service life: The service life of the mask system is dependent on the intensity of usage,
maintenance, and environmental conditions to which the mask is used or stored. As this
mask system and its components are modular in nature, it is recommended that the user
maintain and inspect it on a regular basis, and replace the mask system or any
components if deemed necessary or according to the 'Visual criteria for product inspection'
in the 'Warning' section of this guide.
Note: The manufacturer reserves the right to change these specifications without notice.
AirFit N30i
Pillows
4 to 25 cm H
O
2
0.5 cm H
O
2
1.7 cm H
O
2
21 dBA
14 dBA
25 dBA
18 dBA
Pressure
Flow
(L/min)
(cm H
O)
2
4
20
9
31
15
41
20
49
25
55
AirFit P30i
Pillows
4 to 25 cm H
O
2
0.4 cm H
O
2
1.5 cm H
O
2
21 dBA
14 dBA
24 dBA
17 dBA
English
7