Assembly And Installation; Working; External Connections - Velp Scientifica OHS 200 DIGITAL Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

➢ Assembly and installation

Unpacking
Check the integrity of the unit after unpacking.
The box includes
OHS (XXX) digital overhead stirrer
Power supply cable
Instruction manual
Allen key and allen head screw
Shaft support overhead stirrer
Installation
Place the unit on non-flammable surface
Fix the shaft support to the instrument with the allen head screw
Fasten the unit to the support rod (A00000366 or A00000369) using the double clamp (A00001301)
Secure the receptacle using the ribbon clamp (A00001302)
Slide the stirring shaft into the chuck and tight it
Make sure that the rating value of the instrument corresponds to the one of the power supply
Ensure that the socket provided, with grounding, is compliant with the current safety norms and easy to reach.

➢ Working

Commissioning
Stirring
Timer
Gear (only for
OHS 200)
Torque
Lock

➢ External Connections

USB
Switch on the instrument through the main switch (J)
Display (A) shows software version, last set point value and OFF (if "Mode" is set on Stop, see chapter 0).
Set the operating speed range by clicking Gear key (D) (Only OHS 200).
Adjust speed set point by turning the speed control knob (F).
Click the knob to start stirring.
Speed increases until set point achievement.
A microprocessor ensures constant speed even when the viscosity changes (counter-reaction).
Switch off the stirring by clicking the knob.
Click Timer key (D or H). HH:MM appears on display.
Rotate speed control knob (F) to set the desired value. Click to confirm.
Once timer is set, timer LED lights up:
-
(G) for OHS 200
-
(E) for other models
The LED lights continuously until the instrument starts to stir. When the instrument stirs, the LED flashes.
It's possible to visualize (or modify) countdown in each moment during the analysis by clicking timer key.
When countdown finishes, "End" is visualized on the display.
Set the operating speed range by clicking Gear button (D).
The gear set has its relative LED (C or E) lighted on.
It's possible to select between two gears:
Gear I: Low speed (6 – 400 rpm) High torque (till 200 Ncm).
-
Gear II: High speed (30 – 2000 rpm) Low torque (till 40 Ncm).
-
The torque indicator bar (B), made by four LEDs, shows the intensity of the torque applied
Holding the Lock key (H) for 3 seconds it's possible to lock the control panel.
Unlock the control panel by holding key H for 3 seconds.
When the instrument is locked,
-
LED (I) lights on for OHS 200
-
LED (G) and (I) light on for other models
Each model has USB connection for software upgrading by PC. Software version, dedicated program to be
installed on the PC and software installation guide must be requested by e-mail to service@velp.it
6

Publicité

Table des Matières
loading

Table des Matières