4. REPLACEMENT PROCEDURES

4.1. REPLACING THE WHEELS OF THE MOBILE STAND

The wheels of the Mobile Stand are already adjusted in factory for proper vertically of the column.
In case of the replacement or adjustment of the wheels follow the procedure below:
1
Lift the Mobile Stand about 4/5 cm from the floor,
suspended on wooden support.
2
Remove the wheel by unscrewing the wheel nut
(a) with special 30 mm open wrench.
Record the position of nut (b) on the old wheel.
3
Screw the nut (b) on the new wheel ensuring the
same position of the nut.
In this way the level adjustment of the stand shall
be maintained.
4
Screw thoroughly the new wheel into the pedestal, by the special 30 mm open wrench on wheel nut (a).
Remove the wooden support and put the Mobile Stand on the floor.
5
Distance D shall not exceed 108 mm to ensure the proper resistance to stress of the wheel (see
figure).
F-4
NOTE
WARNING
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
CORRECTIVE MAINTENANCE
mm 102 min.
D
mm 108 max.
Gendex Dental Systems
(b)
(a)