Télécharger Imprimer la page

Maintenance Of The Wall Support - GENDEX Oralix AC Manuel De Service

Masquer les pouces Voir aussi pour Oralix AC:

Publicité

Les langues disponibles

Les langues disponibles

ORALIX AC SYSTEM

4.2.4. Maintenance of the Wall Support

For the maintenance of the Wall Support follow the
procedure below:
Remove the cover (c) and check that the Wall
1
Support is firmly connected to the wall by four
fixation points (the applied torque shall be 10 N/m).
Check also that it is properly levelled.
2
Check the power connections at the terminal block
(d) and inspect the protective earth connection (e)
3
at the ground point, checking for damage or corrosion
which could alter the electrical continuity.
E-6
4519-190-70943_4
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
PREVENTIVE MAINTENANCE
(d)
Gendex Dental Systems
(c)
(e)

Publicité