Télécharger Imprimer la page

Functional Test On Timer Densomat - GENDEX Oralix AC Manuel De Service

Masquer les pouces Voir aussi pour Oralix AC:

Publicité

Les langues disponibles

Les langues disponibles

SETTING TO WORK

6.2. FUNCTIONAL TEST ON TIMER DENSOMAT

1)
Position Tubehead with Collimator toward radiation-safe area, away from
personnel, and cover distal end with 2mm lead.
2)
Switch on the Densomat by pressing the LINE ON switch; verify that the
LEDs light ON, as shown by figure 1.
3)
Take exposure button and move to a radiation safe-area.
4)
Press exposure button and verify that the "X-ray ON" indicator (see figure
2) of the Tubehead 1 should light and X-ray buzzer signal is audible during
exposure.
5)
Verify that audible and visible X-ray ON indicators go OFF when exposure
terminates.
6)
After the exposure any further X-ray command shall be inhibited for the
relevant wait time; if a new exposure is commanded during the wait time
the indication "
"Dead-man" function test
7)
Press "S+" button (see figure 3) until the indication "2.5" appears on dis-
play.
8)
Press and immediately release exposure button, verifying that audible
and visible X-ray ON indicators go OFF immediately after exposure button
is released. The display flashes.
Check mechanical stability of the electrical connections
9)
Move Tubehead, Folding Arm and Extension Arm assembly through all
positions. Periodically press exposure button and verify that audible and
visible X-ray ON indicators operate properly.
10)
Turn power OFF.
11)
Perform Acceptance tests as described in section D.
12)
Complete the installation as described in section B "Closing the unit".
13)
Repeat all the tests from point 2 to 11 for the second Tubehead, if installed.
C-8
" appears on the display.
4519-190-70943_4
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
ORALIX AC SYSTEM
TUBEHEAD 1 SELECTED
DISPLAY
MEDIUM PATIENT
BITEWING
Figure 1
TUBEHEAD1:
X-RAY ON INDICATOR
Figure 2
"S+" BUTTON
Figure 3
Gendex Dental Systems

Publicité

loading