*Bands may not be available in all regions.
FCC ID: 2ACHL-AIR114G
IC: 9103A-AIR114G
The AirSense 11 device complies with FCC Rules and Industry Canada rules. Operation is subject to the following two conditions:
1. This device may not cause harmful interference; and
2. This device must accept any interference received, including interference that may cause undesired operation.
The AirSense 11 device should be installed and operated with minimum distance of 0.59" (15 mm) between the equipment and the
user's body.
Additional information regarding the FCC Rules and IC compliance for this device can be found on ResMed.com/downloads/devices
In Canada, the device has been designed to comply with safety standards to radio waves (SAR) in accordance to RSS-102.
Humidifier
Maximum heater plate temperature:
Temperature cut-out (heater):
Maximum gas temperature (at mask)
1
The air flow for breathing produced by this therapy device can be higher than the temperature of the room. Under extreme
ambient temperature conditions (104ºF/40ºC) the device remains safe.
Humidifier performance
SlimLine/Standard tubing
Mask Pressure
Nominal % RH
cm H
O (hPa)
when delivered to
2
mask at 72ºF (22ºC)
Setting 4 (default
setting)
4
80%
10
80%
20
80%
Climate Control Auto - ClimateLineAir 11
Mask Pressure
Nominal % RH delivered to mask, tested in 72ºF
cm H
O (hPa)
(22ºC) ambient
2
4
85%
10
85%
20
85%
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated
3
Humidifier performance meets ISO 80601-2-74:2017 performance > 12 mg/L BTPS tested at 59°F to 95°F (15°C to 35°C)
Air tubing
ClimateLineAir 11 temperature range
ClimateLineAir 11 temperature cut out
Maximum recommended pressure
Maximum working temperature, when used with a
humidifier
Material
Inner diameter
Length
24
154ºF (68ºC)
165ºF (74ºC)
1
:
106ºF (41ºC)
Nominal % RH delivered
to mask, tested at 72ºF
(22ºC) ambient
Setting 8 (maximum
setting)
100%
100%
100%
ClimateLineAir 11
60 to 86ºF (16 to 30ºC)
106ºF ( 41ºC)
30 cm H
-
Flexible plastic and electrical
components
0.6" (15 mm)
6'6" (2.0 m)
Nominal system output mg/L AH
Setting 4 (default
setting)
6
6
6
Nominal system output mg/L AH
12
12
12
O (30 hPa)
2
1
2
, BTPS
Setting 8
(maximum
3
setting)
12
>12
>12
1
2
, BTPS
SlimLine/ Standard
-
-
30 cm H
O (30 hPa)
2
106ºF ( 41ºC)
Flexible plastic
SlimLine: 0.6" (15 mm)
Standard: 0.74" (19 mm)
SlimLine: 6' (1.8 m)
Standard: 6"6' (2.0 m)